Bauer B-DETETCTION PLUS m Mobile version for temporary measurement of: CO, CO², O², absolute humidity and VOCs (VOCs = volatile organic compounds)
Standard scope of delivery: B-DETECTION PLUS m in a handy case Pressure reduction for inlet pressure B-Detection system Gas sampling unit for direct measurement from the breathing air cylinder, compressor or intake air Dew point sensor incl. drying capsule, PID sensor, rechargeable battery,
SYSTEM LIMITS Maximum system pressure 350 bar Maximum permissible compressor delivery rate 850 l/min Permissible operating temperature +5°C... +45°C
MEASUREMENT PRINCIPLE CO2 Non-dispersive infrared spectroscopy (NDIR) CO, O2 electrochemical Absolute humidity above dew point VOC (residual oil) Photoionization detector (PID) MEASURED VARIABLES AND MEASUREMENT ACCURACY 3 CO2 0... 2,000 ppm ± 2.5 % of the measuring range end value CO 0 ...25 ppm ± 2 % of the measuring range end value O2 0 % - 40 % ± 1 % of the measuring range end value Dew point -70...+60°C or 10 ... 40,000 ppm ± 2°C or 1 ppm + 20% of the display value VOC (residual oil) 0... 1 ppm (related to calibration gas isobutene)
FURTHER TECHNICAL DATA Integrated operating voltage: 24 VDC; stand-alone; 100-250 VAC, 50/60 Hz Flow rate 1.0 ... 3.0 l/min Weight 8.9 kg Dimensions W × H × D [mm] 464 x 366 x 180 SYSTEM TEST/CALIBRATION Visual and leak test once a month Function check at least every 3 months with test gas System check once a year Calibration of dew point sensor: every 24 months